WS8-2NG**
产品详细
|
WS8-10 Standard NG
|
WS8-2 Standard NG
|
|
|
Available Measurement Range
|
330 – 1180 nm
|
330 – 1180 nm
|
|
Absolute (and Other) Accuracies
|
||
|
330 – 375 nm
|
20 MHz 1)
|
10 MHz 2)
|
|
375 – 800 nm
|
10 MHz 1)
|
2 MHz 2)
|
|
800 – 1180 nm
|
8 MHz 1)
|
2 MHz 2)
|
|
Quick coupling accuracy (with multi mode fiber)
|
100 MHz
|
100 MHz
|
|
Wavelength deviation sensitivity/Measurement resolution 3)
|
0.2MHz
|
0.2MHz 6)
|
|
Measurement Speed
|
1000 Hz
|
1000 Hz
|
|
Required Minimum Input Energy and Power 4)
|
0.08 – 60 μJ or μW
|
0.08 – 60 μJ or μW
|
|
FSR of the Fizeau Interferometers (Fine/Wide Mode) 5)
|
2 GHz/20 GHz
|
2 GHz/20 GHz
|
|
Calibration
|
SLR-780 or any other well known laser source Δν < 2 MHz |
I2 stabilized HeNe or any well known laser source Δν < 1 MHz |
|
Recommended Calibration Period |
≤ 1 hour |
≤ 2 minutes
|
|
Warm-up Time |
> 30 minutes |
> 30 minutes |
|
Dimensions |
340 × 252 × 106 mm |
340 × 252 × 106 mm |
|
Weight |
6.4 kg |
6.4 kg |
|
Interface |
USB 2.0 |
USB 2.0 |
1) ± 200 nm around calibration wavelength; outside of this range the accuracy as WS7-30.
2) ± 2 nm around calibration wavelength; outside of this range the accuracy as WS8-10; note 1) also applies.
3) Standard deviation. WS6-200 and higher models require singlemode or photonic crystal fibers to reach this resolution.
4) The CW power interpretation in [μW] compares to an exposure of 1s (generally the energy needs to be divided by the exposure time to obtain the required power).
5) Each instrument in each mode can measure lasers with a linewidth up to 30 % of the correspondig FSR.
6) 100 kHz for special ranges on request.