WS8-2NG**

产品详细

 

 
WS8-10 Standard NG
WS8-2 Standard NG
Available Measurement Range
330 – 1180 nm
330 – 1180 nm
Absolute (and Other) Accuracies
   
330 – 375 nm
20 MHz 1)
10 MHz 2)
375 – 800 nm
10 MHz 1)
2 MHz 2)
800 – 1180 nm
8 MHz 1)
2 MHz 2)
Quick coupling accuracy (with multi mode fiber)
100 MHz
100 MHz
Wavelength deviation sensitivity/Measurement resolution 3)
0.2MHz
0.2MHz 6)
Measurement Speed
1000 Hz
1000 Hz
Required Minimum Input Energy and Power 4)
0.08 – 60 μJ or μW
0.08 – 60 μJ or μW
FSR of the Fizeau Interferometers (Fine/Wide Mode) 5)
2 GHz/20 GHz
2 GHz/20 GHz
Calibration

SLR-780 or any other

well known laser source Δν < 2 MHz

I2 stabilized HeNe or any well known laser source  Δν < 1 MHz

Recommended Calibration Period

≤ 1 hour

 

≤ 2 minutes

 

Warm-up Time

> 30 minutes

> 30 minutes

Dimensions

340 × 252 × 106 mm

340 × 252 × 106 mm

Weight

6.4 kg

6.4 kg

Interface

USB 2.0

USB 2.0

 

1) ± 200 nm around calibration wavelength; outside of this range the accuracy as WS7-30.

2) ± 2 nm around calibration wavelength; outside of this range the accuracy as WS8-10; note 1) also applies.

3) Standard deviation. WS6-200 and higher models require singlemode or photonic crystal fibers to reach this resolution.

4) The CW power interpretation in [μW] compares to an exposure of 1s (generally the energy needs to be divided by the exposure time to obtain the required power).

5)  Each instrument in each mode can measure lasers with a linewidth up to 30 % of the correspondig FSR.

6) 100 kHz for special ranges on request.

 

 

 